AvianGroupUSA logo Avian Technologies LLC
P.O. Box 716
Sunapee, NH 03782-0716
603-526-2420 (office), 603-526-2729 (fax)
MCRL Logo

Goniometric (spectral)
  GSP-1B
  GCMS-3B
  GSP-2
  GCMS-4

Goniometric (photometric)
  GP-5
  GP-200
  GP-700

Color
  CMS-35SP
  CMS-35FS
  DOT-3C

Gloss and Reflectance
  GM-26PRO
  GM-26PRO/Auto
  GM-3D
  RM-3D
  RM-9
  GMX-102
  GMX-203

Haze and Clarity
  HM-150
  HM-65W
  HM-150L2
  HM-150N
  HR-100
  HSP-150
  TM-1D
  HMS-3

Online Systems
  Glossmeter GR-2
  Glossmeter GR-4
  Color CMTS-5000
  Color CMTS-5000T
  Hazemeter HMS-3
  Transmittance TM-100M


Lumenaire
  NL-500


The Murakami Team
Murakami Products
Avian Technologies LLC
Murakami Corporate


HR-100

Haze & Reflectance meter

The HR-100 is designed to measure total light transmittance, diffuse light transmittance, and haze. It is ideally suited to make these measurements for transmissive materials such as plastics, films, glass, and liquids, as well as light reflective materials with a high degree of surface gloss such as metals, coating and mirrors.

The HR-100 conforms to many national and international haze standards, including:

  • IS0 14782 Determination of haze for transparent materials
  • IS0 13468-1 Determination of the total luminous transmittance of transparent materials
  • ASTM D1003 Haze and Luminous Transmittance of Transparent Plastics (D1003 is approximated)

Applications

  • Measuring the total luminous transmittance and haze of plastics, glass, film and liquid
  • Measuring the total reflectance and diffuse reflectance of plated surface and mirror

Specifications

Measurement

1-point measurement & up to 3-point measurement at option.

Spectral Characteristics

Approximating to the illuminant D65 and spectral luminous efficiency for the CIE standard photometric observer.

Measurement range

0 to 50% for haze measurement

Repeatability

Within 0.1% SD(Standard plate H20 measured 30 times) average haze value

Stability

Within ±0.1% SD (standard plate H20 measured for 1 hour)

Measurement cycle

 3-point measurement per second

Production report

Graph on haze is shown on LCD display and can be printed out.

Alarm

When the upper or lower limit of Haze value is beyond the tolerance, alarm signal is given. PC expansion board (option)

Analog output

 4 to 20mA(3CH) optional

Light source

Halogen lamp 12V/100W Life: 2,000hrs nominal.

Receptor

Silicon photodiode

Power Requirement

100V ±5V AC, 50/60 Hz

It is recommended that the instrument power split from motor circuit is used.

Power Consumption

Approx.150W

Communication cable

Optical fiber cable or cable & wire

Conveyance speed of Film

Max 100 m/min.

© 2017 Avian Technologies LLC, all rights reserved.