The HM-150 measures haze using a double-beam configuration, allowing very
high accuracy and good repeatability of measurment. The lights source is halogen, but near-D65 illumination can be realized by the automatic insertion of daylight filter into the optical pathway. Without the filter, the halogen bulb simulates CIE illuminant A. A PC is used for the calculation and instrument control for automatic and easy measurement. The functions, including all haze calculations and automatic averaging, are provided with the included software. |
![]() |
|||||||
|
The HM-150 Hazemeter is compliant with haze transmittance standards ISO 13468, JIS K 7361 ISO 14782, JIS K 7136 (ASTM D 1003, JIS K 7105) |
||||||||
![]() Schematic of the Measurement System |
1 Light Source Lamp 2 Two Branch Fiber 3 Reference Shutter 4 Measurement Shutter 5 Convertible Switch to D65 & A Illuminant 6 Lens 7 Entrance Port Aperture 8 Exit Port Aperture 9 White Plate 10 Detector 11 Reference Fiber 12 Integrating Sphere 13 Light Trap Box 14 Sample |
|
||||||
As with all Murakami instruments, integrated control and data processing software is included with the purchase of the HM-150. |
||||||||